TUBE-TO-TUBE CONNECTOR
IRRIGATION BAG SUPPORT ROD AND HANDPIECE FIXED SUPPORT
24
25
26
27
Place the connector in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C ±2°C, for at least 10
minutes;
Remove the connector from the ultrasonic
tank, brush it with the brush with soft
nylon bristles, and rinse it with distilled
water to eliminate all residues of detergent;
Clean the surfaces of the rod and of the
handpiece holder with a soft cloth with low
ÀEHUUHOHDVHGDPSHQHGZLWKDQHQ]\PDWLF
detergent solution;
Remove the residues of detergent with a
VRIWFORWKZLWKORZÀEHUUHOHDVHGDPSHQHG
with distilled water;
06 CLEANING OF THE ACCESSORIES