Alarms and Troubleshooting
40
For the alarm message ERR 00X, X indicates a character. The specific alarm message may be ERR 001,
ERR 002, and etc.
For all problems related to the potentiometer, judge whether the hardware is faulty. Access the AD value on
the Maintenance screen, move the potentiometer, and observe whether the AD value changes to judge
whether the sensor and related wires are damaged.
Access the AD value displayed on the screen, press the pressure sensor, and observe whether the AD value
displayed changes to judge whether the pressure sensor and related wires are damaged.
9.5 Check after repair
Please refer to appendix