SAFETY
Meets the requirements for double insulation to IEC61010-1
Installation Category III***, up to 230 V to earth and 400 Volts
phase to phase, without the need for separately fused test leads.
If required, fused test leads are available as an optional
accessory.
Complies with the relevant parts of EN 61557:1997-02 as
detailed below.
*** Relates to the transient overvoltages likely to be met in fixed
wiring installations.
E.M.C. In accordance with IEC61326 -1
Operational inaccuracies: Refer to www.megger.com
ENVIRONMENTAL
PROTECTION
IP40 - The instrument is designed for indoor use, or outdoor
use if suitably protected.
TEMPERATURE RANGE
Operating -5˚C to +40˚C up to 90% RH
Storage -25˚C to +65˚C up to 90% RH
GENERAL
Dimensions 245 mm x 200 mm x 95 mm
Weight 1,35 kg with battery
Cleaning Wipe the disconnected instrument with a
clean cloth dampened with soapy water
or Isopropyl Alcohol (IPA).
IEC 61557 / EN 61557
Complies with the following parts of EN 61557, Electrical safety
in low voltage systems up to 1000 V a.c. and 1500 V d.c. -
Equipment for testing, measuring or monitoring of protective
measures:-
Part 1 - General requirements
Part 2 - Insulation resistance
Part 3 - Loop resistance
Part 4 - Resistance of earth connection and equipotential
bonding (Continuity testing)
Part 5 - Resistance to earth
Part 6 - Residual current devices (RCDs)
Specification (Contd.)
34