www.megger.com
DLRO10HD and DLRO10HDX
20
Tests with Automatic Save (DLROHD10X)
Continuous Test
AUTO
3s
2.5
Ω
250
mΩ
µA
100
2.5
1mA
10mA
100
mA
1A
1A
10A
10A
250
Ω
25
Ω
2.5
Ω
250
25
2.5
Tip: Use Clamp test leads (optional accessory).
1. Press
(acknowledged by a beep sound).
Save is available until the memory has 200 test records.
2. Press
TEST
. Test starts (LED shows red).
TEST
3. Arrow direction animated (based on current direction)
and plateau.
4. Continuous measurements are taken on the connected
test piece.
Tests are made every three seconds.
Last test record is shown while the next test is in
progress.
5. Press
TEST
to stop the test.
Test results are saved and their memory slot number is
shown. The Save screen is shown for three seconds.