EasyManua.ls Logo

Megger MIT2500 - Storing Insulation Test Results

Megger MIT2500
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.megger.comMIT400/2
24
G
G
0
0.01
0.1
110
100
W
8
W
M
s
Note: At 30 seconds the measurement is logged internally. This value is not stored in memory unless the "SAVE" button is pressed
AFTER the DAR measurement is complete. See section 9.3.3.
- Figure 26
At the end of the test period the MIT will make a second measurement and display the results as a ratio of the 1st and 2nd measured
values, as below:
G
0
0.01
0.1
110
100
W
8
Note: Second measurement is logged at 0 seconds. Again, this is not stored in memory unless the "SAVE" button is pressed at the
completion of the test, see section 9.3.4
- Figure 27
9.3.2 Polarisation Index
The same method as in 9.3.1 above is used. Only the test times change.
9.3.3 Timed tests:
Timed tests run a countdown timer and make a measurement at the end of the timed period.
1. Select one of the test voltages on the MW measurement mode using the range knob
MW
.
2. Connect the test leads to the circuit to be measured.
Press the DAR-PI-T button on the keypad until the display show "t" on the small digital readout, as below:
G
0
0.01
0.1
110
100
1
W
8
W
M
- Figure 28
3. Press the TEST button. A count down from the limit configured In the SETUP menu will start. At the end of the count down,
the MIT will make a measurement and display the result as a resistance.

Table of Contents

Other manuals for Megger MIT2500

Related product manuals