Safety Warnings
The following Safety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use.
Only use test leads and accessories supplied or approved by Megger Instruments Limited
At any time the G symbol or F symbol is displayed, the user guide and warnings documentation must be consulted to identify the nature of the hazard and any actions
necessary to avoid the hazard
Do not use the instrument if there are any signs of damage
This instrument meets the EMC requirements of Class A applications. Not for use in domestic installations
All test leads, probes and clips must be in good order, clean and with no broken or cracked insulation
Probes and clips should be held behind the finger guard
Test leads not used during a measurement should be disconnected from the Appliance tester
During testing, ensure no hazard will exist as a result of normal running or under fault conditions
During testing the unit under test (appliance) should not be touched, other than using the appropriate accessories, as faulty appliances can present a shock hazard
Do not touch the exposed parts of test leads during tests as hazardous voltages may be present
Do not intentionally connect test leads to live systems or hazardous voltages
Do not touch the IEC extension lead socket pins especially during a test, as hazardous voltages may be present due to a potentially faulty appliance
Do not touch the exposed earth pins of the 230 V test socket during a test, as voltages may be present due to a potentially faulty appliance
Serviceable fuses should only be replaced with those that are suitably rated
Replacement fuses must be of the correct rating and type. Refer to page 33
If this instrument is used in a manner not specified in the supplied documentation, the protection provided by the instrument may be compromised
PAT150
For safety, only connect the PAT to a supply that is properly earthed. If in doubt, the supply should be checked by a qualified electrician
Only perform a mains powered leakage test after the Earth bond and insulation tests have been completed, as this test operates at mains voltage
During mains powered leakage tests the appliance under test will operate. Make sure the appliance is safely secured to ensure no damage or danger is possible
A yearly calibration is recommended with interim checks on measurement accuracy to ensure no equipment can be left in a hazardous live condition through incorrect readings
Only use a Megger approved PAT100 charger. Other chargers may present a fire risk
Do not connect the battery charger to the PAT150R whilst running a test
During testing make sure that the shutter covers the battery charger port. There is a risk of electrocution from exposed terminals. Do not touch any exposed terminals or probe tips
during test
Always remove the mains plug test lead from the mains supply AND the instrument when not in use

Other manuals for Megger PAT120 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Megger PAT120 Series and is the answer not in the manual?

Megger PAT120 Series Specifications

General IconGeneral
BrandMegger
ModelPAT120 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals