www.megger.com
10
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Input power
6 x IEC LR6 1.5 V alkaline (AA)
6 x IEC HR6 1.2 V NiMH rechargeable (AA)
Battery life
1000 TTR tests on a single charge
Storage: 1+ years NiMH, 5+ years alkaline
Battery charging
USB-C when set to NiMH batteries
Protection against alkaline charging
PowerEx PRO NiMH battery charging: 0 to 45 °C.
Output
Voltage Single phase, 1-50 V
Current 0.1 mA – 1 A
Regulatory
Safety IEC 61010-1:2010 + AMD1:2016
EMI/EMC IEC 61326-1:2012
RoHS2 EN50581
Vibe/Shock MIL-STD -810G
Ingress IP54
Transformer testing standards
IEEE C57.152-2013
IEC 60076-1:2011
AS/NZS 6076 1:2014
CIGRE 445 2011
GOST 3484.1-88
Dimensions
22.8 x 10.5 x 7.5 cm 8.98 x 4.1 x 2.95 in
Weight
1 kg 2.2 lbs
Case
Heavy duty over-molded case with built-in connection for hook
strap. Carry case with quick start guide, belt loop hook, and
pouches for included lead set and accessories.
Internal/external data storage
Up to 10 custom vector storage
Up to 10 000 sets of single phase results internal storage
Transferable via USB 2.0 drive
USB-C connection to PC (EXP ONLY)
Communication/control software
USB Interface for PC download with custom GUI
Display
Full colour 88 mm (3.5 in) 320 x 240 px Hi-bright LCD screen with
'auto dim' and 'auto off' to preserve battery life
Printer (optional)
51 mm (2in) thermal printer
Prints all measurement data displayed on GUI
Environmental
Operating -20 ° to 50 °C (-4 ° to +122 °F)
Storage -30 ° to 70 °C (-22 ° to +158 °F)
Relative humidity 0-90 %, non-condensing