EasyManua.ls Logo

Megger TTRU1 - Technical Specifications

Megger TTRU1
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.megger.com
11
Max Technical Specifications
Max Technical Specifications
TTR
Turns ratio measurement
methods Single phase step up
Single phase step down
Turns ratio range
and accuracy
Step down excitation
25-50 V
±0.05 % 0.8 – 1000
±0.10 % 1001 – 2000
±0.30 % 2001 – 15000
±1.0 % 15000 +
1-24 V
±0.10 % 0.8 – 1000
±0.20 % 1001 – 2000
±0.60 % 2001 – 15000
±2.0 % 15000 +
Step up measurement
25-250 V
±0.05 % 0.8 – 200
(most Power Tx)
1-24 V
±0.10 % 0.8 – 200
Excitation current
resolution
Resolution 0.1 mA, 0.1 mA – 100 mA
1.0 mA, 101 mA – 1000 mA
Excitation current
accuracy ±1 % Reading, ±0.1 mA
Frequency accuracy ±1 % Reading, ±0.1 Hz
Phase range 0 – 360 °
Phase accuracy ±0.05 °
Max voltage output 45 V AC peak
SCI
Impedance measurement
methods Single phase
Frequency range 40 – 480 Hz
Impedance measurement
range 0.1 Ω - 700 Ω
Impedance accuracy ±1 % reading, ±0.01 %
±1 % reading, ±0.10 m Ω
Reactance measurement
range 0.1 Ω - 700 Ω
Reactance accuracy ±1 % reading, ±0.01 %
±1 % reading, ±0.10 mΩ
Inductance accuracy ±1 % reading, ±10 µH
Power factor Range 0.1 % – 100 %
Power factor accuracy ±5 % reading, ±0.1 %
AC current accuracy ±0.2 % reading, ±0.1 mA

Related product manuals