31 / 42 9742146
Par.
Nr.
Configuration
options
Use as
Value
range
Unit
Note
224
Actuation mode
final rinse pump
Parameters
0 .. 4
-
Definition: activation of final rinse pump:
0 – final rinse pump = 0; no signal
1 – final rinse pump; activate according to
calculated running time
2 – final rinse pump = booster pump;
energizing as booster pump
3 – final rinse pump = wash pump; energizing
as wash pump
4 free
225
Actuation mode
detergent pump
Parameters
0 .. 4
Definition: activation of detergent pump:
0 – detergent pump; no signal
1 – detergent pump; energizing according to
calculated running time
2 – detergent pump = booster pump, energize
as pressure increasing pump
3 – detergent pump = wash pump;
Activate as wash pump
4 – option detergent pump using negative
pressure dosing
228
Water softener
present?
Parameters
0/1
-
In case of incorporated water softener set to 1
240
Load factory
settings for
configuration data
Parameters
0/1
-
Effective only upon power supply reset
ON/OFF.
Caution! All changes to service parameters
are reset.
Power supply reset must be carried out within
5 minutes, otherwise factory settings will not
be loaded.
Without the power supply reset, information
123 will be displayed.
241
A
0
value
Parameters
0 …60
-
Only in connection with disinfection machine
nos. 5-9 in parameter 201
321
Final rinse pump
output
Parameters
0.1 ...10
L/h
Final rinse pump
Output definition
322
Detergent pump
output
Parameters
0.1 ...20
L/h
Detergent pump
Output definition
326
Rinse aid vent
time
Parameters
0 ... 255
Sec.
Activate the dosing pump for rinse aid
temporarily to remove air from the pipe
327
Detergent vent
time
Parameters
0 ... 100
Sec.
Activate the dosing pump for detergent
temporarily to remove air from the pipe
347
Disinfection
temperature
Parameters
10 ...80
°C/°F
Only in connection with disinfection machine
nos. 5-9 in parameter 201
348
Disinfection
holding time
Parameters
0 ...900
Sec.
Only in connection with disinfection machine
nos. 5-9 in parameter 201