16.2 Error messages and troubleshooting
Figure 3: fault display
Error messages will disappear automatically after the fault has been rectified.
Error messages (extract)
ERR.-no.
Description
Possible cause
001
EEPROM plug-in fault
EEPROM not available/incorrectly
plugged in/defective
Empty or incorrect EEPROM
Replace EEPROM with correct
parameter set
111
Bottom pan leakage
Leak inside the appliance
Pump sump/motor/etc.
Defective leakage detector
Repair error, remove water
117
Door not locked
The pin of the lifting magnet is not
correctly in the locking device.
The magnetic coil of the lifting magnet is
damaged.
Door locking query is not correct.
201
Level not reached during 1st
filling
Fresh water inlet insufficient (water tap
closed)
Aquastop hose kinked
Inlet filter soiled
Aquastop defective
Boiler switch defective
202
Level not reached early enough
during filling
See 201
203
No change detected by the level
switch during emptying
Boost pump defective
Booster pump plug connector loose
Start capacitor defective
Plug connector loose
Plug connector defective
No DSP signal on - from I/O circuit
board
No signal boiler full - from I/O circuit
board
Check boost pump/S2 using manual
control
204
Still no change detected by the
level switch at the end of the
rinse time
See 203
205
Temperature increase not
reached
Boiler heating defective/heating element
thermal fuse
Temperature sensor defective, incorrect
installation position
Boiler protection defective, performance
switch loose
No signal from I/O circuit board