12 Attachment: Technical Information
12.1 Technical Specifications SyncFire 1200
Dimensions: [W x H x D] 43.46 x 4.29 x 61.49 cm (17.11 x 1.69 x 24.21 in)
Mounting Depth Rack: 620 mm
Height Unit Rack: 1 U
19 inch Rackmount: Yes
Mounting Cable Depth: 200 mm (1000 mm rack recommended)
Weight: 10 kg (22.05 lb)
Electrical data (hot-plug power supply unit)
Rated Line Voltage: 100 to 120 V AC (200 to 240 V AC)
Frequency: 50/60 Hz
Effective Power: 500 W
Max. Thermal energy: (for 500 W power supply:)
2088 kJ/hr (1979 BTU/hr) at 100 V AC
2016 kJ/hr (1911 BTU/hr) at 200 V AC
2073 kJ/hr (1965 BTU/hr) at 240 V AC
Power Supply Configuration: 2 x hot-plug power supply for redundancy
Standard Operating
Temperature: 10
◦
to 35
◦
C (50
◦
to 95
◦
F) at sea level with an altitude derating of 1.0
◦
C
per every 305 m (1.8
◦
F per every 1000 ft) above sea level to a maximum of
3050 m (10,000 ft). Maximum rate of change is 20
◦
C/hr (36
◦
F/hr).
Non Operating
Temperature: -30
◦
to 60
◦
C (-22
◦
to 140
◦
F). Maximum rate of change is 20
◦
C/hr (36
◦
F/hr).
Relative Humidity: Operating
10 % ... 85 % (non condensing)
Non-operating
5 to 95 % relative humidity (Rh), 38.7
◦
C (101.7
◦
F)
maximum wet bulb temperature, non-condensing.
Altitude: Operating
3050 m (10,000 ft). This value may be limited by the type and number of options
installed. Maximum allowable altitude change rate is 457 m/min (1500 ft/min).
Non-operating
9144 m (30,000 ft). Maximum allowable altitude change rate is
457 m/min (1500 ft/min).
Source: Hewlett Packard documentation: HPE-ProLiant-DL325-Gen10
42 Date: 31st March 2021 SyncFire 1200