SERVICE - UIS page 22
20. Replacing the door seal
Dismantle the closure cover (see Section 19.3)
Raise the radiation shield and remove the old seal.
(on Model 600-800 release the screws on the radiation shield)
Cut the new door seal at the ends according to the old seal and slide it on the radiation shield.
Fit the radiation shield and press it down.
(On Model 600-800 fit the radiation shield (left with contact strip) and screw it down.)
21. Changing the working chamber seal
Incubators Ixx have two seals
After opening the door the seals can be pulled out and the new seal inserted.
The pre-cut length of the seal should be pushed into the interspace without tension and can be
shortened a little if necessary.
Seal cross-section
Closure cover
Screw
Radiation shield
Seal cross-section
Working chamber seal Working chamber seal
Seal for
inner door
Incubator Ixx
Oven Uxx/Sxx