30 D24030 | Date 11/2018
Operation and control
Interior lighting (only for models with light module)
Adjustment range: 0 to 100 % in steps of 1%
LIGHT
%
40
The interior light is only active at temperatures of up to
40 °C. If this temperature is exceeded, the interior light
switches off automatically. The light display will then show
“Temp too high”.
LIGHT
%
0
Temp too high
5.4.5 Operation with digital backwards counter with target time setting,
adjustable from 1 minute to 99 days ( Timer)
In timer operation, you can adjust the time the appliance runs at the set value. The appliance
has to be in manual operating mode for this.
1. Press the activation key to the left of
the timer display. The timer display is
activated.
TIMER
-
Ende
9:00 23.11.
--h- m
2. Turn the turn control until the desired
duration is displayed – in this example 4
hours 30 minutes. The approximate end
time is shown beneath, in a smaller font.
TIMER
End
0
13:30 23.11.
04 mh 3
Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes)
format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum
duration adjustable is 99 days 00 hours.
3. Press the confirmation key to confirm.
The display now shows the remaining time
in a large font and the approximate end time
in a smaller font beneath. The status display
shows „Timer active“.
TIMER
30m04h
Ende
13:30 23.11.
Timer active
12.Sept.2012
13:44
4. Now, as described under 5.4.2 , set the individual values for temperature, air flap position
etc. which you want the appliance to operate at. The set values can be changed at any
time during the timer. The changes are effective immediately.