36 D24030 | Date 11/2018
Operation and control
Save the selection by pressing the confirma-
tion key. The min setting (undertemperature
protection) is automatically activated.
ALARM
max
auto
40
0.0K
min
15.0°C .0°C
+
-
1. By turning the turn control, adjust the
desired lower alarm limit value, in the
example on the right 35.5 °C.
The lower alarm limit value cannot be set
higher than the top one. If no undertem-
perature protection limit is required, set
the lowest temperature.
ALARM
0.0K
min
°C
auto
max
°C
35 5.
40 0.
+
-
2. Press the confirmation key to confirm.
The max display (overtemperature protec-
tion) is activated.
ALARM
0.0
K
min
°C
auto
max
°C
35 5.
40 0.
+
-
3. By turning the turn control, adjust the
desired upper alarm limit value, in the
example on the right 38.5 °C.
The monitoring temperature must be set
sufficiently high above the maximum set
temperature. We recommend 1 to 3 K.
ALARM
0.0
K
min
°C
auto
max
°C35.5 38 5.
+
-
4. Accept the upper alarm limit value by
pressing the confirmation key. The setting
of the automatic temperature monitor
(ASF) is automatically activated (
auto).
ALARM
max
38.5°C
0.0K
min
35.5°C
auto
+
-
5. With the turn control, select ON (
) or
OFF ().
ALARM
max
38.5°C
0.0K
min
35.5°C
auto
+
-
6. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
38.5°C35 5.
0.0
+
-