23 03/2021
Operation and control
In the following example, the temperature calibration is performed for the example temperature of 30°C:
Make 2-point calibration under Appliance Settings on
Press to return to the home screen
CAL 1 Temp
DEV 1 +/-
DEV 2 +/-CAL 2 Temp
00.0
°C
0.0
K
0.0
K
00.0
°C
Select CAL1Temp to set the compensation temperature CAL1
Scroll vertically to set the compensation temperature CAL1 to 30°C
Press to conrm your selection
Scroll vertically to set the compensation correction value to 0.0 K
Press to conrm your selection
Press to return to appliance settings
Press to return to the home screen
0
°C
30 .
CAL 1
0
K
+ 0.
DEV 1