24 03/2021
Operation and control
memmert
31.6 °C
Position the sensor of a calibrated reference instrument centrally in the appliance's
working chamber
Wait until the appliance reaches the set temperature and displays 30 °C.
The reference instrument displays for example 31.6 °C
Set compensation correction value for Cal1 to +1.6K (measured actual value minus
setpoint)
Press to conrm your selection
After the calibration procedure, the temperature measured by the reference instru-
ment will now also be 30 °C.
30.0 °C
31.6 °C
6
K
+ 1.
DEV 1
30.0 °C
30,0 °C