25 03/2021
Malfunctions, warning and error messages
6. Malfunctions, warning and error messages
Description Cause Action
70.9 °C
E-113
Actual temperature has exceeded the set
alarm temperature
■ Appliance must cool down until the
actual temperature falls below the
set alarm temperature
■ Increase alarm temperature if nec-
essary
70.9 °C
E-130
The PTC was triggered. Appliance is too
hot, the water level in the appliance may
have been too low during operation
■ Switch off appliance
■ Allow to cool down
■ Monitor water level
■ Restart appliance
■ Contact Service
if this occurs frequently
70.9 °C
E-134
Temperature sensor error
■ Restart appliance
■ Contact Service
if this occurs frequently
70.9 °C
E-111
Software error
■ Restart appliance
■ Contact Service
if this occurs frequently
70.9 °C
E-132
Memory access error
■ Restart appliance
■ Contact Service
if this occurs frequently
70.9 °C
E-135
CRC error (internal memory error)
■ Restart appliance
■ Contact Service
if this occurs frequently