25 03/2021
Malfunctions, warning and error messages
6. Malfunctions, warning and error messages
Description Cause Action
70.9 °C
E-113
Actual temperature has exceeded the set
alarm temperature
Appliance must cool down until the
actual temperature falls below the
set alarm temperature
Increase alarm temperature if nec-
essary
70.9 °C
E-130
The PTC was triggered. Appliance is too
hot, the water level in the appliance may
have been too low during operation
Switch off appliance
Allow to cool down
Monitor water level
Restart appliance
Contact Service
if this occurs frequently
70.9 °C
E-134
Temperature sensor error
Restart appliance
Contact Service
if this occurs frequently
70.9 °C
E-111
Software error
Restart appliance
Contact Service
if this occurs frequently
70.9 °C
E-132
Memory access error
Restart appliance
Contact Service
if this occurs frequently
70.9 °C
E-135
CRC error (internal memory error)
Restart appliance
Contact Service
if this occurs frequently