Operation and control
5.4 Outlet valve
The drain valve can become overheated while emptying the water bath. Touching its
surface can cause burns. Use heat-resistant protective gloves to turn the drain valve
on/off or let the tempering medium cool down rst and empty the water bath when it
has reached a lukewarm temperature.
NOTICE
Observe the following points before emptying the water bath:
Allow tempering medium to cool down before draining
Outlet valve can overheat during operation, please observe safety instructions
Open outlet valve slowly to prevent water from leaking out uncontrolled
When emptying the water bath, let the tempering medium drain either directly into a
suitable and adequately sized container or push a suitable tube onto the outlet valve
(outlet valve diameter: 12 mm), which hooks into a container.
5.5 Calibration
The appliances are temperature calibrated and adjusted at the factory. In case read-
justment should be necessary later on for example due to inuence of the chamber
load – the appliance can be calibrated customer-specically using two calibration tem-
peratures of your choice:
Cal1 Temperature calibration at low temperature
Cal2 Temperature comparison at high temperature
To guarantee perfect control, we recommend calibrating the appliance once a year.
CAL 1
+1,6°C
CAL 2
-0,4°C
Factory calibration
50°C 100°C
0°C
30°C
WARNING