EN | Preparation for use
12
5. Preparation for use
5.1. Transport and storage
Device with cover: Thedeviceisshippedintwopolystyrenetraysfastenedtogetherinacardboardbox.
Built-in version:Thedeviceisshippedinacardboardboxwithbubblewrap.
Thisoriginalpackagingmustbeusedforanyonwardandreturntransport.Thedevicemustalwaysbetransportedandstoredinanuprightposition.Thedevice
must be transported to the installation site in a completely packed state. After unpacking the device, check for completeness and possible transport damage.
Transport and storage temperature 0 °C – 70 °C
Transport and storage humidity limitation max. 80%
5.2. Installation requirements
Operating temperature 10 °C – 40 °C
Humidity limit max. 70%
above sea level ≤3000m
5.2.1. Assembly variants
Installation in the treatment unit
Thedirectintegrationofthewaterdisinfectionsystem intothetreatment unitshouldbe the
preferred solution, as the hose and compressed air routes to the device can be kept as short
as possible.
In the unit, vibration-free suspension of the device on load-bearing parts of the treatment unit
must be ensured.
Installation with cover
If it is not possible to install the device directly in the treatment unit, the set-up can also be
carried out in the optionally available cover.
The cover should be mounted on the connection box of the treatment unit. The external display
and a transformer are already integrated in the cover.