6.4. Error messages
Assembly, modications and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety
instructions)!Forfurtherinformationandassistanceincarryingoutrepairs,retrotting,faultanalyses,etc.,theMETASYStechnical
customer service is also available!
Error message Possible cause Countermeasures Consequence
LED 1 lights up green
LED3lightsupyellow
Buzzer tone
Disinfectionoftheservicewateris
not carried out
Change the GREEN&CLEAN WK
bottle(see7.3)!Ifthebottlecannot
be changed immediately, press the
RESET button to deactivate the
buzzer.
The device remains functional, the
dentist‘sworkowisnotinterrupted,
but no disinfection takes place.
However,ashortbuzzerwillcontinue
to sound periodically.
Iftheerrormessagewasonlycaused
by an air bubble, the error message
willdisappearonitsownatthenext
attempt.
RESET
LEUCHTEN
BLINKEN
1
RESET
2
RESET
3
RESET
1
2 3
4
LED 1 lights up green
LED2ashesred
Buzzer sound
Level probe in the mixing container is
dirty,ornowatersupply!
Firstcheckwhetherthemainwater
taporthewatersupplylineis
open. Clean or replace the mixing
containerprobe(see7.2.2).
Allvalvesandthepumpareswitched
off.Thebuzzercanbeswitchedoff
withtheRESETbutton.Oncethe
causehasbeenrectied,theWEK
canonlybereactivatedbyswitchingit
off and on again.
RESET
RESET
RESET
LEUCHTEN
BLINKEN
1
RESET
LED 1 lights up green
LED 2 lights up red
Buzzer sound
Defective double membrane pump,
oroverow!
Cleantheoverowsensor.Check
double membrane pump and circuit
board(see7.2.3).
Allvalvesandthepumpareswitched
off.Thebuzzercanbeswitchedoff
withtheRESETbutton.Oncethe
causehasbeenrectied,theWEK
canonlybereactivatedbyswitchingit
off and on again.
RESET
RESET
RESET
LEUCHTEN
BLINKEN
1
RESET
2
RESET