AquaLab TDL 12 TROUBLESHOOTING
SOLUTION:
1. On the measurement screen, toggle to the correct model us-
ing the up and down arrow keys. The moisture content value
updates to correspond with the model you select.
2. If the correct model is not available, the model may not be
loaded on the instrument.
(a) To determine which models are loaded on the instrument,
cycle to the Configurtation tab, select %Moisture and then
the loaded models appear.
3. If the correct model is not available, load the appropriate model
using AquaLink 4 Software. The AquaLab TDL can hold a
total of 100 models at any one time. You may need to re-
move a model using the Software or use the delete option in
the %Moisture Calibration menu before you can add a new
one. Any model that you remove from the instrument with
AquaLink 4 stores in the software to use later.
12. PROBLEM:
Moisture Content displayed is not correct.
SOLUTION:
1. Model selected may not be correct for the product being tested.
(a) Toggle through the available models to find a more appro-
priate model.
(b) If the model is correct but not giving correct moisture con-
tent values it may be necessary to generate a new model
for the product or update an existing model. For infor-
mation about generating a model, contact METER for
updating a model.
13. PROBLEM:
Moisture content does not show up on the screen.
SOLUTION: