MI 3103 GigaOhm Specifications
19
6.2 General specifications
Power supply 6 V DC (4 × 1.5 V alkaline battery IEC LR14)
Typical operating time >70 h
Auto power off yes, approximately 10 min of no activity
Dimensions (w × h × d) 280 x 70 x 80 mm
Weight (without accessories) 490 g
Display custom LCD
Protection classification Class II (double insulation)
Over-voltage category CATIII 300 V
Pollution degree 2
Protection degree IP 40
Reference conditions
Reference temperature range 10
C 30 C
Reference humidity range 40 %RH
70 %RH
Operating conditions
Working temperature range 0
C 40 C
Maximum relative humidity 95 %RH (0
C 40 C), non-condensing
Storage conditions
Temperature range -10
C +70 C
Maximum relative humidity 90 %RH (-10
C +40 C)
80 %RH (40
C 60 C)
Accuracies apply for 1 year in reference conditions. Additional error in operating
conditions is 1 % of measured value and 1 digit.
Fuses
Instrument protection M 0.315 A / 250 V, 5
22 mm