Calibrateur multifonctions CX 1651
160 Manuel de l'utilisateur
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: RJUNction (?) <DNPD>
: TYPE (?) <CPD> { B | E | J | K | N | R | S | T }
: PRT
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?) <DNPD>
: TYPE (?) <CPD> { PT385 | PT392 | NI }
: NRESistance (?) <DNPD>
MEASure subsystem
This subsystem allows to control the internal multimeter of CX 1651 calibrator. It sets the multimeter’s functions
and reads measured values.
Keyword Parameters
MEASure
?
: CONFigure
: VOLTage
: CURRent
: MVOLTage
: RESistance
: FREQuency
: TEMPerature
: RTD
: TYPE (?) <CPD> { PT385 | PT392 }
: NRESistance (?) <DNPD>
: THERmocouple
: TYPE (?) <CPD> { B | E | J | K | N | R | S | T }
: RJUNction (?) <DNPD>
: SGS
: VOLTage (?) <DNPD>
: OFF
TESTer subsystem
This subsystem allows to control features of CX 1651 calibrator when used as tester. It starts selected test
procedure and reads its result.
Keyword Parameters
TESTer
: RUN <DNPD>
: RESUlt ?
Operation complete
*OPC <cr>
This command sets the OPC bit in the ESR (Event Status Register) when all pending operations are complete.