7 Troubleshooting
904 Titrando
■■■■■■■■
35
Problem Cause Remedy
■ Select the user-defined titration rate and
decrease the control range.
■ Select the user-defined titration rate and
increase the maximum rate.
■ Select the user-defined titration rate and
increase the minimum rate.
The results are
spread widely.
The minimum dosing rate is
too high.
Select user-defined titration rate and decrease
the minimum rate (see manual/help of the
software used).
The electrode responds too
slowly.
Replace the electrode.