13. TECHNICAL SPECIFICATIONS
TYPE
SpiderView digital Holter ECG and high resolution ECG recorder.
USA
only
This device is a AECG device type 1.
NUMBER OF LEADS
Depending on the cable used, the device records the following leads:
2 or 3 bipolar leads
12 leads* (6 peripheral leads + 6 precordial leads)
3 leads XYZ to calculate 12 leads according to Dower matrix*
* Available with SyneScope Holter software.
ELECTRICAL CHARACTERISTICS
Standard quality frequency re-
sponse:
0.05 - 25 Hz
High-quality and uncompressed
mode frequency response
0.05 - 80 Hz
Common-mode rejection: 94 dB
Input dynamic range: ± 16 mV
Power supply: One 1.5 V AA long-life lithium battery (for seven 24-hour
recordings) or one NIMH type rechargeable batteries with
a minimum capacity of 1500 mAh.
MECHANICAL CHARACTERISTICS
Weight 85g (3oz) without battery
(25g (0.9oz) for standard 1.5V AA)
Dimensions 97 x 54 x 23 mm (3.8’’ x 2.1’’ x 0.9’’)
For information on operating temperature, storage temperature, atmospheric pressure
range, humidity, read conditions for use and storage on paragraph 4.3.
ACQUISITION CHARACTERISTICS
Storage on flash cards
Acquisition frequency 1000 Hz
Resolution 10 µV (16 bits)
Pacemaker spike detection
High-resolution ECG 1000 Hz, 2.5 µV