CHAPTER II System Function and Configuration
P 21
II.1.C.e
Insulation Calibration
Operation Procedure
(a) Operation is the same before, in the subordinate menu of calibration, using
[↑] or [↓] to select (5) Insu.Cal. LCD display as following
(b) Press [Enter], LCD display as following
(c) Press S1[UP] or S2[DOWN] to adjust code, let the reading value match or
closer to the standard voltage as much as possiable
(d) After setup finished, press S5 [DONE] to save
(e) If you want to abandon data, press S6[QUIT] to exit from this screen