Areas of application
43
Programme
selection
Special programmes matched to the mobile injector units are stored
for reprocessing ophthalmological instruments. Disinfection is carried
out thermally.
Mobile injector unit
A 204
The mobile injector unit A 204 is divided into 2 levels; it has a spray
arm and may only be used with the Ophthalmic programme.
The top level is fitted with various connections for reprocessing
lumened instruments, such as rinsing and suction hand-pieces, and
cannulae.
The lower level of the mobile injector unit is designed to take inserts
and mesh trays for reprocessing instruments without lumen.
Mobile injector unit
A 207
The mobile injector unit A 207 has 3 levels with 2 spray arms and may
only be used with the OphthaTraysA207 programme.
An injector manifold with silicone hoses with Luer Lock connectors is
located on the upper level. Trays and mesh trays for ophthalmological
theatre sets with integrated injector manifolds can be connected to
this.
The 2 lower levels are designed to take inserts and mesh trays for
reprocessing instruments without lumen.