Settings
16
Q-POD conguration
This app allows a user to congure all parameters specic to Q-PODs. These are unique to the
Q-POD that is being used to input the values.
To duplicate Q-POD parameters, the action(s) should be repeated on all other Q-PODs.
Q-POD Name
This can be personalized. Click on the text box and input up to a maximum of 8 characters.
Screen brightness
Adjust to the preferred brightness from 1 to 7 using the arrows or click on the box to access
the keypad and type-in the value.
Sound volume
Each Q-POD can emit a sound when alerts/alarms are triggered. This can be activated or deactivated
using the slider button. The sound can be adjusted to the desired volume by using the arrows or
click on the box to access the keypad and type-in the value. By default, the sound is deactivated.
Flowmeter offset
The owmeter has been calibrated in order to achieve volume precision of +/- 2%. Laboratory
glassware accuracy can vary a lot. This setting provides a user with the ability to adapt the
owmeter to their current glassware accuracy by using an offset function.
Adjust the offset by using the arrows or click on the box to access the keypad and type in the
value. To gain back the original calibration settings, set the value back to 0.
Accessories
Water sensor
A water sensor can be connected to the system unit to stop
the system from producing water in case of water spillage.
Up to 3 sensors can be connected in series to cover a wider
surface area.
Foot pedal
The pedal is connected to the base of the Q-POD.
In the Q-POD conguration app, activate the foot pedal via
the slider button.
Dispense using the foot pedal
1. Press once and release to dispense in full ow.
2. Press and hold to start dispensing in drop-by-drop, keep
pressing to increase the ow rate up until full ow is reached.
Release at the chosen ow rate.
3. Press once to stop the ow dispensing.
Feed water conductivity sensor
A conductivity sensor can be installed inside the system
to monitor the feed water quality entering the system. If
bad water quality is detected, an alarm that stops water
production will appear, informing the user that the system
is at risk and to check the feed water quality. This will
protect the downstream purication technologies in case of a
problem.