Quick start
6
Note: When multiple Q-PODs are installed and dispensing is taking place on a Q-POD, the
other Q-PODs will display the following information:
Dispensing ongoing
Q-POD 1
Species the Q-POD on which the dispensing
is taking place.
During this time, only the Information menu is accessible. Dispensing functions will become
available again as soon as the dispensing operation is nished on the initial Q-POD.
Dispense report
A dispense report is automatically produced after each dispensing operation. A dispensing
operation is considered to be all dispenses with less than 10 second intervals between them. This
can be interrupted at any time by pressing on the dispense report icon that becomes available
on the home screen as soon as a dispense is initiated.
1. Visualize
Click on the dispense report icon:
2. Personalize (assign a name or experiment number)
In the user ID eld, enter the content of your choice (max 15 characters)
3. Export
- Scan the QR code from a mobile device to instantly retrieve the report.
- Click on the export button to save the report on a USB key.
4. Automatic archiving
The dispense report icon on the home screen will automatically disappear after 5 mins of inactivity.
Each dispense report, whether consulted or not, is automatically archived and can be retrieved
anytime by going in the history report section available in the information menu. Go to page 11
Information > History for further information.