Cleaning the Inlet Strainer
Purpose
ly
Cleaning the Inlet Strainer removes any trapped debris.
The purpose of the Inlet Strainer is to prevent a large particle from entering
the System.
If the Inlet Strainer becomes clogged, then feed water does not flow free
to the System.
When
The Inlet Strainer should be cleaned when the following Alert message is
displayed.
Alert message = EXAMINE INLET STRAINER
u suspect it is clogged.
The Inlet Strainer should also be cleaned whenever yo
Procedure
Fol the t Strainer.
Step Action
low steps below to clean the Inle
1 Go to STANDBY Mode.
2 Shut off the feed water supply.
3 Unscrew the Inlet Strainer from the feed water supply.
4 Detach the tubing on the other end of the Inlet Strainer.
5 Flush water backwards through the Inlet Strainer.
6 Apply 3 to 4 turns of new white tape to the threads of the feed
water pipe.
7 S k onto the feed water pipe. crew the Inlet Strainer bac
8 Attach the tubing to the other end of the Inlet Strainer.
9 Open the feed water supply valve.
10 Go to READY Mode.
Continued on next page
- 91 -

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore Milli-Q Direct 8 System and is the answer not in the manual?

Millipore Milli-Q Direct 8 System Specifications

General IconGeneral
BrandMillipore
ModelMilli-Q Direct 8 System
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals