Initial Zero
select P77
press
"ENTER"
[P77 0]
displayed
press
"1"
press
"ENTER"
[P77 1]
displayed
press
"ENTER"
[oo ####]
displayed, wait (time dependent upon the value of P30)
[od 0.000]
displayed
press
"ENTER"
[oC ####]
displayed, New Zero Count, zero calibration complete
Direct Zero Entry
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial zero at that time.
select P67
press "ENTER"
[P67 0] displayed
press "1"
[P67 1] displayed
press "ENTER"
[Pc 0.] displayed e.g. value ’0’ after a P99 reset
enter the recorded zero reference of previous zero (#####)
press "ENTER"
[oC #####] displayed, new zero count, entry complete