Initial Zero
select P77
press
"ENTER"
[P77 0]
displayed
press
"1"
press
"ENTER"
[P77 1]
displayed
press
"ENTER"
[oo ####]
displayed momentarily then,
[od 0.000]
displayed
press
"ENTER"
[oC ####]
displayed, New Zero Count, zero calibration complete
Direct Zero Entry
Direct entry is intended for use when replacing software or hardware and
it is not convenient to perform an initial zero at that time.
select P67
press "ENTER"
[P6 70] displayed
press "1"
press "ENTER"
[P6 71] displayed
press "ENTER"
[P c0.] displayed e.g. value ‘0’ after a P99 reset
enter the recorded zero reference count of the previous zero (#####)
press "ENTER"
[o C#####] displayed, new zero count, entry complete.
PL-516 55