EasyManuals Logo

Mindray BeneView T8 Service Manual

Mindray BeneView T8
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
5-56
3. Disconnect the cable of the infrared communication board, and release the snap lock to
remove the infrared communication board.
4. Disconnect the cables of the NIBP inflation pump and the relief valves from the
parameter board. Snip off the cable ties to remove the pump. Release the snap locks to
remove the valves.
Snap loc
k

Table of Contents