3-3
Pressure calibration
C.O. test
Mainstream CO
2
test
Leakage test
Performance test
Sidestream and
Microstream
CO
2
tests
Calibration
Leakage test
Performance test
AG tests
Calibration
ICG test
BIS test
RM test
Interconnecting
function
CCO/SvO
2
test
Output calibration
Performance test
NMT test
Sensor check
EEG test
PiCCO test
ScvO
2
test
Nurse call relay performance test
Analog output performance test
If the user suspects that the nurse call or analog
output does not work well.
Electrical Safety Tests
Electrical safety tests Refer to A Electrical Safety Inspection.
Other Tests
Power on test
1. When first installed or reinstalled.
2. Following any maintenance or the replacement of
any main unit parts.
Touchscreen calibration
1. When the touchscreen appears abnormal.
2. After the touchscreen is replaced.
Recorder check Following any repair or replacement of the recorder.
Network print test
1. When first installed.
2. Whenever the printer is serviced or replaced.
Device integration check
1. When first installed.
2. Following any repair or replacement of the
external device.
Battery check Functionality test
1. When first installed.
2. Whenever a battery is replaced.

Table of Contents

Other manuals for Mindray BeneView T8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T8 and is the answer not in the manual?

Mindray BeneView T8 Specifications

General IconGeneral
BrandMindray
ModelBeneView T8
CategoryMonitor
LanguageEnglish

Related product manuals