3-15
Block the gas inlet of the module or watertrap. The sidestream
CO
2
flowrate is slower than 10ml/min and an alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
The displayed CO
2
value is within 6±0.05.
Microstream CO
2
test
Block the gas inlet of the module or watertrap. An alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
The displayed CO
2
value is within 6±0.05.
AG test
When AG flowrate is slower than 10ml/min, an alarm of AG
Airway Occluded is given. It indicates that there is no leakage.
The fan inside the AG module works properly.
The measurement accuracy of CO
2
, N
2
O, O
2
and AA (AA
represents an anaesthetic agent) meets the product specifications
in the Operator’s Manual.