6-28          BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 
6.8 Disassembling the MPM Module 
1.  Removing the front panel assembly 
  As shown in the figure, use a contact wrench to remove the screw for the back end terminal of 
the module; use a small flat-bladed screwdriver to remove the module retention tabs. Use a 
Phillips screwdriver to remove the front panel screw. Then, the front panel of module can be 
removed. 
 
 
 
 
  As shown in the following figure, loosen and remove the screw between the front panel and the 
bracket, and unplug the air tube at the air nozzle. Then the front panel can be removed. 
 
2.  Remove the parameter board 
  As shown in the following figure, loosen and remove the board screws. Remove    pull out the pump 
and valve cable and NIBP air tube. Then, the parameter board can be removed. 
 
   
Remove the 
bracket screw 
 
Use a contact wrench to 
remove the terminal screw 
 
Remove the module 
retention tabs 
Remove the screw 
 
Parameter board screws and 
cable 
 
Remove along this direction