EasyManuals Logo

Mindray DC-45 User Manual

Mindray DC-45
303 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #286 background imageLoading...
Page #286 background image
B-2 Electrical Safety Inspection
B.2 Device Enclosure and Accessories
B.2.1 Visual Inspection
Test Item Acceptance Criteria
The enclosure and
accessories
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
B.2.2 Contextual Inspection
Test Item Acceptance Criteria
The enclosure and
accessories
No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells,
particularly from ventilation holes).
No taped notes that may suggest device deficiencies or
operator concerns.
B.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
îš„ Main unit label
îš„ Integrated warning labels
B.4 Protective Earth Resistance
 Plug the probes of the analyzer into the device’s protective earth terminal and protective
earth terminal of the AC power cord.
îš„ Test the earth resistance with a current of 25 A.
îš„ Verify the resistance is less than limits.
LIMITS
ALL COUNTRIES R = 0.2 Ω Maximum
B.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
The following outlet conditions apply when performing the Earth Leakage test:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-45 and is the answer not in the manual?

Mindray DC-45 Specifications

General IconGeneral
BrandMindray
ModelDC-45
CategoryMedical Equipment
LanguageEnglish

Related product manuals