9-36 Structure and Assembly/Disassembly
2. Unscrew eight M4 X 12 screws on the control panel assembly.
3. Lift the upper cover up around the margin. Disconnect all plugs on the upper cover assembly.
The control panel assembly and the main unit fall apart then.
4. Unscrew the M3 X 8 combination screw on the cast-aluminum base to remove the control panel
assembly.
Press the clips
inwards.