9-44 Structure and Assembly/Disassembly
4. Unscrew the five M4 X 8 screws used to fix cables.
9.3.26 AC Connecting Board
1. Follow steps 1-6 in Chapter 9.3.8 to remove the electronic assembly (including power input
assembly).
2. Unscrew four M3 X 6 screws, disconnect four connecting plugs, and then remove AC
connecting board.