Software Installation &Maintenance 6-7
9. Double-click the "BackToDoppler" tool in the desktop to return to the ultrasound screen.
Navigate to [Setup] > [Maintenance] > [Setup] and click [Disable ms-setting].
10. Select [OK] > [Save] to exit and power off the system.
Note: After clicking [Enable ms-setting], you can make settings for the operating system on
Windows interface. After the settings, click [Disable ms-setting] on the maintenance screen to
disable the function immediately. Otherwise, the performance or operation of the ultrasound system
may be affected.
Phone Activation
If the ultrasound system is not connected to the internet, activate Windows 10 by the phone.
1. Log on to the system with the Service account, go to [Setup] > [Maintenance] > [Setup] >
[Enter Windows], and input the password.
For details about obtaining password, please refer to "6.4 Enter Windows".
2. Input “cmdin the address column in the pop-up Windows Explorer and press [Enter] key.
3. Input slui 4in the DOS prompt and press [Enter] key. Phone activation wizard Menu will be
displayed as below.

Table of Contents

Other manuals for Mindray DC-88

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-88 and is the answer not in the manual?

Mindray DC-88 Specifications

General IconGeneral
BrandMindray
ModelDC-88
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals