Electrical Safety Inspection
A-16
ELECTRICAL SAFETY INSPECTION
8- Patient Auxiliary Current
couldbeusedinstead.
Changeanotherprobetoconfirmifthefailiscausedbyconsole.
Inspectwiringforbadcrimps,poorconnections,ordamage.
If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the
system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
USA:
For ECG Input and ECG Input and transducer
10μA Normal Condition
50μA Single Fault Condition
OTHER COUNTRIES:
For
ECG Input
10μA Normal Condition
50μA Single Fault Condition
For ECG Input
100μA Normal Condition
500μA Single Fault Condition