Electrical Safety Inspection
A-16
ELECTRICAL SAFETY INSPECTION
8- Patient Auxiliary Current
couldbeusedinstead.
Changeanotherprobetoconfirmifthefailiscausedbyconsole.
Inspectwiringforbadcrimps,poorconnections,ordamage.
If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the
system problem. Remove unit from
operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
USA
For ECG Input and ECG Input and transducer
10μA Normal Condition
50μA Single Fault Condition
OTHER COUNTRIES:
For
ECG Input
10μA Normal Condition
50μA Single Fault Condition
For ECG Input
100μA Normal Condition
500μA Single Fault Condition