3-8
Pace pulse
Pace pulses meeting the following conditions are marked by the
PACE indi cator.
Pulse indicator
Amplitude:
Width:
Rise time:
±4 to ±700 mV (3/5-lead)
0.1 to 2 ms
10 to 100 µs
When tested in accordance with the ANSI/AAMI EC13 Sections
4.1.4.1 and 4.1.4.3, the heart rate meter rejects all pulses meeting
the following conditions.
Pulse rejection
Amplitude:
Width:
Rise time:
Min. input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
20 V/s RTI
ST segment measurement
Measurement range -2.0 to +2.0 mV
Precision
-0.8 to +0.8 mV: ±0.02 mV or ±10%, whichever is greater
Beyond this range: Undefined
Update period 10 s
Mortara Software Package
Lead naming style AHA, EURO
Lead fault
The lead resistance is no greater than 51 kΩ and it is in parallel
with a 0.047 µF capacitor, it will not cause a lead fault condition.
For 3/5-lead, differential offsets ≤ ±300 mV, it will not cause a
lead fault condition.
For 12-lead, differential offsets ≤ ±500 mV, it will not cause a
lead fault condition.
Sensitivity selection
1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5),
10 mm/mV (×1), 20 mm/mV (×2) and AUTO
Sweep speed 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth (-3 dB)
Diagnostic mode:
Monitor mode:
Surgery mode:
0.05 to 150 Hz (M08A module)
0.5 to 40 Hz
1 to 20 Hz