5.2 NIBP Calibration
Figure 5-1 NIBP Calibration
Calibration method:
Based on the precision of 50mmHg (6.7kPa), increase the pressure step by step. The maximum
error at any pressure point within the NIBP measurement range of the patient monitor should be
no more than ±3mmHg (±0.4kPa). Decrease the pressure step by step. The maximum error at any
pressure point within the NIBP measurement range of the patient monitor should be no more than
±3mmHg (±0.4kPa).