EasyManuals Logo

Mindray DPM 4 Service Manual

Mindray DPM 4
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
5-3
5.1.6 SpO2
Select PLETH as the HR source of DPM4, and put the finger into the SpO
2
sensor. The screen
should display the PR and SpO
2
values normally. The normal SpO
2
value is above 97%.
5.1.7 IBP
1. Test fixture
Physiological signal simulator
2. Test procedure
IBP1 test:
Set the BP sensitivity of the ECG simulator to 5uv/v/mmHg, BP to 0mmHG, and the IBP
channel 1 to ART. Enter the IBP PRESSURE ZERO menu of the DPM4, zero Channel 1,
and then return to the main screen. Set the BP of the simulator to 200mmHg. Enter the IBP
PRESSURE CALIBRATE menu of the DPM4, conduct calibration, and then exit the IBP
PRESSURE CALIBRATE menu.
Set the BP value of the simulator respectively to 40mmHg, 100mmHg and 200mmHg. Then
the screen of the DPM4 should display 40±1mmHg, 100±2mmHg and 200±4mmHg.
Set the simulator output to ART wave. Then the screen of the DPM4 should display relevant
waveform properly.
Unplug the IBP probe. Then the screen should prompt “IBP: Transducer 1 OFF!” and “IBP:
Transducer 2 OFF!”
Plug the OHMEDA cable to the IBP1 channel. Then the prompting message “IBP:
Transducer 1 OFF!” disappears.
IBP2 test:
Plug the IBP cable to the IBP2 channel, and repeat the procedure in Section .

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals