5-2
6. RESP measurement: Set the baseline impedance to 1K, the respiration impedance to 0.5
and 3, and the respiration rate to 30 and 120. The respiration rate should be 29 – 31, 118
–122.
7. PVC test: Set the simulator to the PVC mode, and set the occurrence times. The relevant
PVCS should be obtained.
8. Set the simulator as follows: RR: 40, baseline impedance: 2K, RESP waveform: 3:1. Open
the apnea alarm, set the respiration resistance to 0, and set various alarm time. Alarms
should be given.
5.1.4 Temperature
1. YSI probe
Select YSI probe from the manufacturer menu, select YSI temperature probe as the test
fixture, set the analog resistance to 1.471K, 1.355K and 1.249K. Then the TEMP parameter
should be 35±0.1, 37±0.1 and 39±0.1.
2. CY-F1 probe
Select CY-F1 probe from the manufacturer menu, select CY-F1 temperature probe as the
test fixture, set the analog resistance to 6.534K, 6.018K and 5.548K. Then the TEMP
parameter should be 35±0.1, 37±0.1 and 39±0.1.
5.1.5 NIBP
Connect the NIBP simulator, adult cuff and accessories, and then connect the module CUFF and
clockwise screw it tightly.
1. After the simulator self-test, press <ENT> to enter the ADULT analog blood pressure mode.
Set the blood pressure to the 255/195/215 mmHg level, SHIFT to +15, and the HR to
80BPM. Set DPM4 to the adult mode. Press <START>. Then the results will be obtained in
about 30s. The measured results should be respectively 270±8mmHg, 210±8mmHg and
230±8mmHg.
2. Press <ESC> and <> on the simulator to enter the NEONATE mode. Set the blood
pressure to the 120/80/90 mmHg level, HR to 120bmp, and DPM4 to the pediatric mode.
Press <START>. Then the results will be obtained in about 30s. The measured results
should be respectively 120±8mmHg, 80±8mmHg and 90±8mmHg.
3. Press <ESC> and <> on the simulator to enter the NEONATE mode. Set the blood
pressure to the 60/30/40 mmHg level, SHIFT to -20, HR to 120bmp, and DPM4 to the
neonate mode. Change the simulator accessory to the neonatal cuff. Press <START>. Then
the results will be obtained in about 30s. The measured results should be respectively
40±8mmHg, 10±8mmHg and 20±8mmHg.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals