3-11
When tested in accordance with the ANSI/AAMI EC13 Sections
4.1.4.1 and 4.1.4.3, the heart rate meter rejects all pulses meeting
the following conditions.
Pulse rejection
Amplitude:
Width:
Rise time:
Min. input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
20 V/s RTI
ST segment measurement
Measurement range -2.0 to +2.0 mV
Precision
-0.8 to +0.8 mV: ±0.02 mV or ±10%, whichever is greater
Beyond this range: Undefined
Update period Updated every 16 valid beats