3-14
Masimo SpO
2
Module
All SpO
2
sensors specified in the section Masimo SpO2 Accessories meets the following
specifications when used with Masimo SpO
2
module.
SpO2
Measurement range 1 to 100%
Resolution 1%
Precision
70 to 100%:
70 to 100%:
70 to 100%:
0% to 69%:
±2% (adult/pediatric, non-motion conditions)
±3% (neonate, non-motion conditions)
±3% (in motion conditions)
Undefined.
Refreshing rate 1 s
Averaging time 2-4 s、4-6 s、8 s、10 s、12 s、14 s、16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion accuracy ±2%
PR
Measurement range 25 to 240 bpm
Resolution 1 bpm
Precision
±3 bpm (non-motion conditions)
±5 bpm (in motion conditions)
Refreshing rate 1 s