3-20
* Precision applies for breath rates of up to 80 bpm. For breath rates above 80 bpm, accuracy
complies with EN ISO 21647/ISO 21647/ISO 9918 (4 mmHg or ±12% of reading whichever is
greater) for EtCO2 values exceeding 18 mmHg. To achieve the specified accuracies for breath
rates above 60 breaths/minute, the Microstream® FilterLine H Set for Infant/Neonatal (p/n
006324) must be used.The accuracy specification is maintained to within 4% of the values
indicated in the above table in the presence of interfering gases according to EN ISO
21647/ISO 21647 Section Eleven, Part 101.