Part II General Information
Function and Performance Check 220
6.3 General Check
6.3.1 Check Flow
Check the control panel
Check the monitor
Check peripherals
Check I/O port
Check ECG
6.3.2 Check Items
6.3.2.1 Control Panel
Procedure Acceptance Criteria
1. Check the buttons, keys and knobs.
Perform the check from left to right and from up down.
All keys and knobs are effective.
2. Check the functionality of the trackball.
Press Freeze key to enter the frozen state and move
the trackball.
Press Measure to enter the measurement state.
Measure the vertical distance and horizontal
distance or perform other trackball operations.
The trackball works smoothly and
responds quickly to cursor movement.
The rotation direction of the trackball is
consistent with the movement direction of
the cursor.
6.3.2.2 Display
Procedure Acceptance Criteria
Adjust the brightness of the display. Brightness adjustment
Auto adjustment: Choose Setup > System > General >
DISPLAY and click Brightness auto adjustment.
Adjust the contrast of the display. Choose Setup > System > General > DISPLAY and
click the plus sign to increase the contrast or the
minus sign to decrease the contrast.

Table of Contents

Other manuals for Mindray MX7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray MX7 and is the answer not in the manual?

Mindray MX7 Specifications

General IconGeneral
Display Size15.6 inches
Transducer Ports3
PortabilityPortable
TypeUltrasound System
Display TypeLCD
Battery Life120 min
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler
ConnectivityWi-Fi, Ethernet, USB
ApplicationsVascular, Cardiology, OB/GYN, Urology
Probe Connectors3
Operating SystemWindows

Related product manuals