Part III Appendix
Description of Self-Test Items 244
Troubleshooting suggestion
a If the test result is Error, restart the equipment and re-perform the self-test. If the problem
persists, recover the equipment.
b If the test result is FAIL, replace the main board.
2.14 Z0212 FPGA Temperature Test
Upper-level test item
Communication Test Between PC Module and MF FPGA
Test content
The PC reads the FPGA core temperature through the PCIe bus.
Test failure analysis
If the test result is Error, the drive malfunctions.
If the test result is FAIL, the FPGA core temperature is too high.
Troubleshooting suggestion
a If the test result is Error, restart the equipment and re-perform the self-test. If the problem
persists, recover the equipment.
b If the test result is FAIL, check whether the fans are blocked, whether the FPGA radiator is
properly assembled, and whether the heat-conducting cream and radiator are in good contact.
2.15 Z0301 Fan Speed Test
Upper-level test item
Communication Monitoring Test
Test content
Read the fan speed through the UART of the SM ARM and judge whether the current fan speed
meets the requirements.
Test failure analysis
If the test result is Error, the drive malfunctions.
If the test result is FAIL, the fans are blocked (do not rotate), or a fault is detected on the main
board.
Troubleshooting suggestion
a If the test result is Error, restart the equipment and re-perform the self-test. If the problem
persists, recover the equipment.
b If the test result is FAIL, check whether the fans rotate properly. If so, replace the main board.
2.16 Z0302 Fan Speed Adjustment Test
Upper-level test item
Communication Monitoring Test
Test content

Table of Contents

Other manuals for Mindray MX7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray MX7 and is the answer not in the manual?

Mindray MX7 Specifications

General IconGeneral
Display Size15.6 inches
Transducer Ports3
PortabilityPortable
TypeUltrasound System
Display TypeLCD
Battery Life120 min
Imaging ModesB, M, Color Doppler, Power Doppler, Pulsed Wave Doppler
ConnectivityWi-Fi, Ethernet, USB
ApplicationsVascular, Cardiology, OB/GYN, Urology
Probe Connectors3
Operating SystemWindows

Related product manuals